Chinese | English
HOME ABOUT US PRODUCTS FEEDBACK SERVICE CONTACT US
Laboratory Equipment
Microplate Reader
Microplate Washer
Hematology Analyzer
Chemistry Analyzer
Spectrophotometer

Hospital Equipment
ECG
Ventilator
Monitor
Ultrasound Scanner
X-ray Machine
Film Processor
Infant Incubator
Radiant Warmer
Phototherapy Unit
Dental Unit
Hospital Furniture

Veterinary Products
Autoclave
Ultrasound Scanner
Surgical Light
Defibrillator

Products
TW-1100 Infant Radiant Warmer
Standard Composition
¡ñServo Control Mode:air temperature
¡ñ Instant Parameter Display: set temp, skin temp, heat power rate;
¡ñ Audible/Visual Alarms: temp deviation, sensor failure, power failure, temp override;
¡ñ Tilting Bed with Rotary Radiant Head
¡ñ X-ray Cassette Tray under infant bed, examination light of high illuminance;
¡ñ Use anti-blast quartz infrared tube as radiant heating source.
Technical Specification:
This equipment belongs to Class¢ñ, Type BF, continuous operation common equipment.
POWER
Model TW-1000-1E.............................................................................AC 110 / 120 V, 50Hz / 60Hz
Model TW-1000-2E........................................................................... AC 220 / 240V, 50Hz / 60Hz
Input Power..................................................................................................................... ¡Ü1000VA
TEMPERATURE CONTROL RANGE AND RELATIVE SPECIFICATION
Control mode........................................................................................................................baby mode
Temperature control range....................................................................................................25¡æ¡«37¡æ.
Skin temperature sensor measure range..............................................................................20¡æ¡«42¡æ
Deviation between sensor measure temperature and control temperature...................................¡Ü0.5¡æ
Skin Sensor Precision.............................................................................................................. ¡Ü0.3¡æ
Bed surface temperature Uniformity................................................................................................ ¡Ü1¡æ
Temperature Rise Time ............................................................................................................. ¡Ü45min
Bed tiltable.........................................................................................continues adjustable in ¡À10¡ã
ALARM
Sensor failure alarm.....................................................alarm with audible and visual when sensor open                                                                                       circuit, short circuit or disconnect, cut off power.
Over temperature alarm.................................................alarm with audible and visual when displayed                                                                                                                         temperature was approach 38¡æ.
Deviation alarm......................................after temperature equilibrium, alarm with audible and visual                                                                                          when displayed temperature was 1¡æ above or below control temperature, cut off power if above 1¡æ.                                                       
Power failure alarm................................................... alarm with audible and visual when power failure
ENVIRONMENT TEMPERATURE
(It is a suggestion that please do not use outside of the allow environment.)
Operate temperature range.............................................................................................18¡æ¡«30¡æ
Store temperature range...........................................................................................  -20¡æ¡«+55¡æ
ENVIRONMENT HUMIDITY
Operate humidity range........................................................................................ 30%RH¡«75%RH
Store humidity range......................................................................................................... ¡Ü93%RH
ATMOSPHERIC PRESSURE
Transport and Store atmospheric pressure range................................................500hPa¡«1060hPa
Operate atmospheric pressure range...................................................................700hPa¡«1060hPa
STREAM VELOCITY
Ambit environment stream velocity................................................................................... ¡Ü0.3m/s

Tel: 0086-25-58651928   Fax:0086-25-83254278 Email: teriomed@gmail.com
© 2007-2008 teriomed.com all rights reserved. support:Rurui.net