TW-2100 Infant Radiant Warmer
Standard Composition
¡ñTemperature Control Mode: pre-heating control, servo control & manual control;
¡ñ Instant Parameter Display: set temp, skin temp, heat power rate;
¡ñ Built-in APGAR Timer and >37¡æpreset function
¡ñ Audible/Visual Alarms: temp deviation, sensor failure, power failure, temp override;
¡ñ Tilting Bed with Rotary Radiant Head;
¡ñ X-ray Cassette Tray under infant bed, examination light of high illuminance;
¡ñ Use anti-blast quartz infrared tube as radiant heating source.
Technical Specification:
This equipment belongs to Class¢ñ, Type BF, continuous operation common equipment.
POWER Model TW-2000-1E................................................................................AC 110 / 120 V, 50Hz / 60Hz Model TW-2000-2E............................................................................. AC 220 / 240V, 50Hz / 60Hz Input Power....................................................................................................................... ¡Ü1000VA
TEMPERATURE CONTROL RANGE AND RELATIVE SPECIFICATION Control mode..................................................................Pre-warm mode, manual mode, baby mode Temperature control range.......................................................25¡æ¡«37¡æ,for set temperature ¡Ü37¡æ 37¡æ¡«38¡æ,for set temperature >37¡æ Skin temperature sensor measure range..............................................................................20¡æ¡«42¡æ Deviation between sensor measure temperature and control temperature...................................¡Ü0.5¡æ Skin Sensor Precision.............................................................................................................. ¡Ü0.3¡æ Bed surface temperature Uniformity............................................................................................ ¡Ü1¡æ Temperature Rise Time .......................................................................................................... ¡Ü45min Bed tiltable...........................................................................................continues adjustable in ¡À10¡ã
ALARM Sensor failure alarm....................................................alarm with audible and visual when sensor open circuit, short circuit or disconnect, cut off power. Over temperature alarm...........................alarm with audible and visual when displayed temperature was approach 38¡æ for set temperature ¡Ü37¡æ, or approach 39¡æ for set temperature >37¡æ, cut off power. Deviation alarm............................................after temperature equilibrium, alarm with audible and visual when displayed temperature was 1¡æ above or below control temperature, cut off power if above 1¡æ. Power failure alarm............................................. alarm with audible and visual when power failure APGAR time.................................................... alarm with audible when run between 50 seconds to 1 minute, 4 minutes 50 seconds to 5 minutes, 9 minutes 50 seconds to 10 minutes. ENVIRONMENT TEMPERATURE (It is a suggestion that please do not use outside of the allow environment.) Operate temperature range.................................................................................................18¡æ¡«30¡æ Store temperature range.......................................................................................... -20¡æ¡«+55¡æ ENVIRONMENT HUMIDITY Operate humidity range....................................................................................... 30%RH¡«75%RH Store humidity range............................................................................................................ ¡Ü93%RH ATMOSPHERIC PRESSURE Transport and Store atmospheric pressure range..............................................500hPa¡«1060hPa Operate atmospheric pressure range.....................................................................700hPa¡«1060hPa
STREAM VELOCITY Ambit environment stream velocity................................................................................ ¡Ü0.3m/s
|