TI-3100 Infant Incubator
Standard Composition: ¡ñServo Control Mode: air temperature, skin temperature; ¡ñ Instant Parameter Displaying: set temp, air temp ,skin temp, internal humidity, heat power rate; ¡ñ Audible & Visual Alarms: power failure, sensor failure, higher & lower temp, air& skin temp override, fan failure, system failure; ¡ñ >37¡æ Preset Mode: power failure reset memory, alarm silence, keypad lock, self-test function; ¡ñ Hood with dual slope: 6 hand-ports; left & right cabinets, top-shelf; ¡ñ RS-232 Output Connector (Use for network system ).
POWER Model TI-3000-1E............................................................................................110 / 120 Vac, 50Hz / 60Hz Model TI-3000-2E.........................................................................................220 / 240 Vac, 50Hz / 60Hz Power Input ........................................................................................................................... ¡Ü400VA
Refer to IEC60601-1, this equipment belongs to Class¢ñ, Type BF. Its operation is continuous.
ALARM Power Alarm Power Failure.............................................Activates if primary power to the Incubator fails or the power wire is accidentally disconnected from the wall receptacle. Fan Alarm Fan Failure...................................................................................Activates if Fan motor malfunctions. Sensor Alarm Air Sensor Failure.............................................................Air Sensor malfunction, refer unit to service.
Sensor Out of Position....................................Activated when Sensor is not in correct position to perform calibration or continue operation.
Skin Sensor Disconnected........................................Activated during Skin Control Mode when Skin Sensor is disconnected from Temperature Controller. Skin Sensor Failure..............................................................Activated when Skin Sensor malfunctions.
Deviation Alarm High Air Temperature...............................Activated if displayed Air Temperature is 1.5¡æabove set point. Low Air Temperature...............................Activated if displayed Air Temperature is 2.5¡æbelow set point. High Skin Temperature.........................Activated if displayed Skin Temperature is 1.0¡æabove set point. Low Skin Temperature......................Activated if displayed Skin Temperature is 1.0¡æbelow set point.
Over Temperature Alarm Over Air Temperature......................Activated when displayed Air Temperature was approach 38¡æ for set temperature ¡Ü37¡æ, or approach 40¡æ for set temperature >37¡æ in Air Mode. Over Skin Temperature........................Activated when displayed Skin Temperature was approach 38¡æ for set temperature ¡Ü37¡æ, or approach 39¡æ for set temperature >37¡æ in Skin Mode. System Alarm System Failure .................................................Activated if Temperature Control System malfunctions.
TEMPERATURE CONTROL RANGE AND CORRELATIVE SPECIFICATIONS Air Temperature...........20¡æ¡«37¡æ,for set temperature ¡Ü37¡æ, 37¡æ¡«39¡æ,for set temperature >37¡æ Skin Temperature........20¡æ¡«37¡æ,for set temperature ¡Ü37¡æ, 37¡æ¡«38¡æ,for set temperature >37¡æ Temperature Rise Time .......................................................................................................... ¡Ü30min Temperature Variability............................................................................................................. ¡À0.5¡æ Temperature Uniformity.............................................................................................................. ¡Ü1¡æ Skin Sensor Precision................................................................................................................. ¡Ü0.3¡æ HUMIDITY Humidity Display Range............................................................................................ 5%RH¡«99%RH
ENVIRONMENT REQUIREMENT Environment Temperature Operation Temperature Range................................................................................+20¡æ ¡« +30¡æ Store Temperature Range........................................................................................-40¡æ ¡« +55¡æ Environment Humidity Operation Humidity Range.............................................................................................30%¡«75% Store Humidity Range.................................................................................................10%¡«100% Atmospheric Pressure Range............................................................................ 700hPa¡«1060hPa OTHER SPECIFICATION Noise Level Within Hood Environment...................................... ¡Ü55dB(Environment noise below 45dB) Carbon Dioxide (CO2) Level within the Hood..........................................Less than 0.5% when a mixture of CO2 in air is delivered at 750ml/min at a point 10cm above the center of the mattress. Air Velocity Over Mattress....................................................................................less than 0.35m/s Infant Bed Max Load.................................................................................................................10kg Transfusion Shelf Max Load.......................................................................................................2kg Cabinet Max Load .................................................................................................................... 6kg
FUNCTION OF ¡° ¡±SILENCE KEY Make the alarm to be silence for 10 minutes by press this key, following: Deviation Alarm Sensor Alarm Fan Alarm
FACTORY DEFAULT SETTINGS Temperature Control Mode.......................................................................Air Temperature Mode Air Set Temperature.......................................................................................................................32¡æ Skin Set Temperature....................................................................................................................34¡æ Temperature Units..........................................................................................................................¡æ
NOTE: Open Access Panel or Access Door or the use of supplies or other equipment within the Incubator, which can alter the air flow pattern, may effect temperature uniformity, temperature variability, the correlation of the Incubator temperature reading to center mattress temperature and skin temperature.
Optional Part
YC Infant Bed adjustment, adjust tile ¡À10¡ãwith step-less.
SC Double wall canopy.
GL Blue light Phototherapy Unit, LED digital timer, display therapy time and total time.
GB Top shelf for put other device.
HL Protect machine, to prevent hit in moving.
YEC Infant scale, to weighing infant weight.
SGL
Blue light Phototherapy Unit with up and below,
LED digital timer, display therapy time and total time.
Up phototherapy unit used blue tube light source,
below phototherapy unit used blue LED light source.
Technical Specification
Specifications of hood |
TI-3000 |
TI-2000 |
TI-1000 |
TI-800 |
Mattress to Hood |
450 mm |
450 mm |
450 mm |
410 mm |
Observation side of hood |
Two incline |
Two incline |
Two incline |
Two incline |
Operate port |
6 (2 Iris ports) |
6 (2 Iris ports) |
6 (2 Iris ports) |
4 (2 Iris ports) |
Tubing grommets |
4 |
4 |
4 |
2 |
Front access panel |
826¡Á310 mm |
826¡Á310 mm |
826¡Á310 mm |
826¡Á280 mm |
Mattress Tray Size |
630¡Á360 mm |
630¡Á360 mm |
630¡Á360 mm |
630¡Á360 mm |
Technical Specifications: |
|
Power requirement |
¡Ü400VA |
¡Ü400VA |
¡Ü400VA |
¡Ü400VA |
Temperature control mode |
Air and skin servo control |
Air and skin servo control |
Air servo control |
Air servo control |
AirTemp Control Range |
20¡æ¡«37¡æ |
20¡æ¡«37¡æ |
20¡æ¡«37¡æ |
20¡æ¡«37¡æ |
Temperature override mode to |
20¡æ¡«39¡æ |
20¡æ¡«39¡æ |
20¡æ¡«39¡æ |
---- |
SkinTemp Control Range |
30¡æ¡«37¡æ |
30¡æ¡«37¡æ |
---- |
---- |
Temperature override mode to |
30¡æ¡«38¡æ |
30¡æ¡«38¡æ |
---- |
---- |
Skin temp sensor accuracy |
¡Ü0.3¡æ |
¡Ü0.3¡æ |
---- |
---- |
Humidity display range |
5%¡«99% |
---- |
---- |
---- |
Temperature fluctuation |
¡Ü0.5¡æ |
¡Ü0.5¡æ |
¡Ü0.5¡æ |
¡Ü0.5¡æ |
Mattress temperature uniformity |
¡Ü0.8¡æ |
¡Ü0.8¡æ |
¡Ü0.8¡æ |
¡Ü0.8¡æ |
Temperature rise time |
¡Ü30 min |
¡Ü30 min |
¡Ü30 min |
¡Ü30 min |
Noise level within hood environment |
¡Ü55dB |
¡Ü55dB |
¡Ü55dB |
¡Ü55dB |
Alarms |
|
Power failure alarm |
Yes |
Yes |
Yes |
Yes |
Fan failure alarm |
Yes |
Yes |
Yes |
Yes |
Sensor failure alarm |
Yes |
Yes |
Yes |
Yes |
High air temperature alarm |
+1.5¡æ |
+1.5¡æ |
+1.5¡æ |
+1.5¡æ |
Low air temperature alarm |
-2.5¡æ |
-2.5¡æ |
-2.5¡æ |
-2.5¡æ |
High skin temperature alarm |
+1.0¡æ |
+1.0¡æ |
---- |
---- |
Low skin temperature alarm |
-1.0¡æ |
-1.0¡æ |
---- |
---- |
Over air temperature alarm |
¡Ü38.0¡æ; ¡Ü40.0¡æ |
¡Ü38.0¡æ; ¡Ü40.0¡æ |
¡Ü38.0¡æ; ¡Ü40.0¡æ |
¡Ü38.0¡æ |
Over skin temperature alarm |
¡Ü38.0¡æ; ¡Ü39.0¡æ |
¡Ü38.0¡æ; ¡Ü39.0¡æ |
¡Ü38.0¡æ; ¡Ü39.0¡æ |
---- |
System failure alarm |
Yes |
Yes |
Yes |
---- |
Alarm silence function |
Yes |
Yes |
Yes |
---- |
Operational condition |
|
Environment temperature |
+20¡æ¡«+30¡æ |
+20¡æ¡«+30¡æ |
+20¡æ¡«+30¡æ |
+20¡æ¡«+30¡æ |
Environment relative humidity |
5%¡«99% |
5%¡«99% |
5%¡«99% |
5%¡«99% |
Environment air velocity of flow |
<0.3m/s |
<0.3m/s |
<0.3m/s |
<0.3m/s |
Others |
|
RS-232Connector |
Yes |
Yes |
Yes |
----- |
Rail |
Yes |
Yes |
Yes |
----- |
Infant bed tilt adjusting |
¡À10¡ã step-less |
¡À10¡ã step-less |
¡À10¡ã step-less |
----- |
Mattress tray Max. load |
10kg |
10kg |
10kg |
10kg |
Transfusion pole Max. load |
2kg |
2kg |
2kg |
2kg |
Cabinet Max. load |
6kg |
6kg |
6kg |
---- |
Packing size (2 cartons) |
100¡Á58¡Á80 cm |
100¡Á58¡Á80 cm |
100¡Á58¡Á80 cm |
100¡Á58¡Á98 cm |
Gross Weight |
120kg |
120kg |
120kg |
83kg |
|