TI-1100 Infant Incubator
Standard Composition:
¡ñServo Control Mode: air temp; ¡ñ Instant Parameter Displaying: set temp, air temp and heat power rate; ¡ñ Audible & Visual Alarms: power failure, sensor failure, higher & lower temp, air temp override, fan failure, system failure; ¡ñ >37¡æ Preset Mode: power failure reset memory, alarm silence, keypad lock, self-test function; ¡ñ Hood with dual slope: 6 hand-ports; left & right cabinets, top-shelf; ¡ñ RS-232 Output Connector (Use for network system ).
Specification
POWER Model TI-1000-1E..........................................................................................110 / 120 Vac, 50Hz / 60Hz Model TI-1000-2E............................................................................................220 / 240 Vac, 50Hz / 60Hz Power Input ............................................................................................................................... ¡Ü400VA
Refer to IEC60601-1, this equipment belongs to Class¢ñ, Type BF. Its operation is continuous.
ALARM Power Alarm Power Failure.................................................Activates if primary power to the Incubator fails or the power wire is accidentally disconnected from the wall receptacle. Fan Alarm Fan Failure........................................................................................Activates if Fan motor malfunctions. Sensor Alarm Air Sensor Failure..................................................................Air Sensor malfunction, refer unit to service.
Deviation Alarm High Air Temperature...............................Activated if displayed Air Temperature is 1.5¡æabove set point. Low Air Temperature................................Activated if displayed Air Temperature is 2.5¡æbelow set point.
Over Temperature Alarm Over Air Temperature......................................................Activated when displayed Air Temperature was approach 38¡æ for set temperature ¡Ü37¡æ, or approach 40¡æ for set temperature >37¡æ in Air Mode. System Alarm System Failure ................................................Activated if Temperature Control System malfunctions.
TEMPERATURE CONTROL RANGE AND CORRELATIVE SPECIFICATIONS Air Temperature............20¡æ¡«37¡æ,for set temperature ¡Ü37¡æ, 37¡æ¡«39¡æ,for set temperature >37¡æ Temperature Rise Time ............................................................................................................ ¡Ü30min Temperature Variability................................................................................................................ ¡À0.5¡æ Temperature Uniformity................................................................................................................. ¡Ü1¡æ
ENVIRONMENT REQUIREMENT Environment Temperature Operation Temperature Range..................................................................................+20¡æ ¡« +30¡æ Store Temperature Range............................................................................................-40¡æ ¡« +55¡æ Environment Humidity Operation Humidity Range.................................................................................................30%¡«75% Store Humidity Range......................................................................................................10%¡«100% Atmospheric Pressure Range............................................................................... 700hPa¡«1060hPa OTHER SPECIFICATION Noise Level Within Hood Environment.................................... ¡Ü55dB(Environment noise below 45dB) Carbon Dioxide (CO2) Level within the Hood...........................Less than 0.5% when a mixture of CO2 in air is delivered at 750ml/min at a point 10cm above the center of the mattress. Air Velocity Over Mattress............................................................................................less than 0.35m/s Infant Bed Max Load.....................................................................................................................10kg Transfusion Shelf Max Load............................................................................................................2kg Cabinet Max Load .......................................................................................................................... 6kg
FUNCTION OF ¡° ¡±SILENCE KEY
Make the alarm to be silence for 10 minutes by press this key, following: Deviation Alarm Sensor Alarm Fan Alarm
FACTORY DEFAULT SETTINGS Temperature Control Mode.................................................................................Air Temperature Mode Air Set Temperature....................................................................................................................32¡æ Temperature Units...............................................................................................................................¡æ
NOTE: Open Access Panel or Access Door or the use of supplies or other equipment within the Incubator, which can alter the air flow pattern, may effect temperature uniformity, temperature variability, the correlation of the Incubator temperature reading to center mattress temperature and skin temperature. |